Welcome to our site! EDAboard.com is an international Electronics Discussion Forum focused on EDA software, circuits, schematics, books, theory, papers, asic, pld, 8051, DSP, Network, RF, Analog Design, PCB, Service Manuals... and a whole lot more! To participate you need to register. Registration is free. Click here to register now.
I am doing the same project as yours
you will need to buy 75 ohm cable or adapter from minimum loss pad to your fixture
the others cable can use 50ohm
VNA-->50ohm environment-->MLP-->75ohm environement--fixture
50ohm environment
From VNA till MLP, it depends
If your...
ANy pro out there
Can guide me how to do this calibration ?
I don't have the setting for the setup
I am using Agilent PNA-X for Network analyzer
Since i am using self-made cal board, i need to setup my own value and my own setup for this calibration or do i use default value ?
I never use that before
You can find from net for the guide whether does it support any smith chart utilities in it.If yes, then it will make your work easier a lot
Hi all,
I came through this process with some doubts
If i am measuring OIP3 in 75ohm environment , means there will be two different frequency feed into my input.
Does it cause any unbalanced output ?Do i need a balun to balance it ?:-|
To all,
When i setting up for my PNA-x for the measurement, Why do I need to set my IF Bandwidth below 1KHz ?
Is there any specs for it or we defined ourselves ?
All response are appreciated and welcomed
Thank you
thanks but, i already have the minimum loss pad.What i needed is the adapter from output of the MLP to the cal board
The adapter have to be 75ohm nominal impedance, N-type (female) to SMA (male)
hi all,
I am calibrating for 75 ohm environment with matching pad 50-75ohm and 3.5mm 50ohm Rf cable and TRL cal board(75ohm)
Problems starts here
to calibrate off the loss for the matching pad, i need to connect a adapter to my matching pad.
The adapter have to be 75ohm nominal impedance...
I know.My question is regarding the external components like inductor and caps, do i need to change their value when i changing termination impedance from 50ohm to 75ohm ?