Welcome to our site! EDAboard.com is an international Electronics Discussion Forum focused on EDA software, circuits, schematics, books, theory, papers, asic, pld, 8051, DSP, Network, RF, Analog Design, PCB, Service Manuals... and a whole lot more! To participate you need to register. Registration is free. Click here to register now.

i need a notch filter

Status
Not open for further replies.

kakar133

Member level 2
Joined
Sep 16, 2010
Messages
49
Helped
0
Reputation
0
Reaction score
0
Trophy points
1,286
Location
USA
Activity points
1,579
i need a notch filter to remove noise from ECG signal please help me if there is any ic or any circuit diagram of notch filter is available.......
 

I would not recomend you using a simple notch filter becouse it can degrade your signal. Is it jut a homemade project or a a real medical application project? Can you show your schematic here. Maybe your circuit can be improved with out a use of a notch filter as using the right DRL design (i am assuming you are using a DRL). And finally i would recomend you, if possible to implement the notch filter digitally. Any way, if is just a homemade project here is a basic design manual of filters provided by ti: https://focus.ti.com/lit/an/sloa093/sloa093.pdf

At the notch filter i would recomend you to change the R5 and R6 for a pot becouse they are used to tune the center frequency of the filter.

Cya
 
What do you mean by DRL design ?

---------- Post added at 20:41 ---------- Previous post was at 20:40 ----------

The circuit i have implemented to acquire ECG is attached and at the output i get a noisy ECG where i wish to filter it..
Kindly suggest me a filter.
 

 the op07's are they connected as shown?

did you zero the op07's by shorting the in inputs and connect a pot to pins 1 and 8 to zero the opamp like on the spec sheet for the op07's h**p://www.ti.com/lit/ds/symlink/op07c.pdf

did you place a .1mf cap by each ic +v to ground, and to each ic -v to ground if dual voltage?

is it LNA118 did not see the part in digikey so I cant comment on the chip. what are you using for u1?




What do you mean by DRL design ?

---------- Post added at 20:41 ---------- Previous post was at 20:40 ----------

The circuit i have implemented to acquire ECG is attached and at the output i get a noisy ECG where i wish to filter it..
Kindly suggest me a filter.
 