How to test quality of earth connection ?


Post new topic  Reply to topic    EDAboard.com Forum Index -> Electronic Elementary Questions -> How to test quality of earth connection ?
Author Message
steve_rb



Joined: 17 Feb 2004
Posts: 64


Post01 Aug 2004 6:32   

how to test earth connection


I am using a XRF instrument and I want to be sure earth connection has sufficient quality since instrument has a lot of boards and any discharge can cause disaster. Could experts please advice how I can test my instrument earth connection quality with tools like Ohmeter or oscilloscope.

Steve
Back to top
sergio mariotti



Joined: 22 Dec 2003
Posts: 467
Helped: 42
Location: Italy


Post09 Aug 2004 17:21   

how to check earth connection


I think you are speaking about the Heart connection on the mains supply line.
If yes, witout a specific instrument, try this, but remember, the operation contain a risk.

1) I assume Your mains is Line+Neutral.
2) I assume the Neutral is "good enough" to have a potential near to 0 V and enough stable during the time.
3) I assume that a some kind of GND contact on the socket is REALLY connected to the Heart, also if with a bad conection. THIS IS FUNDAMENTAL FOR SEFTY !
4) Disable an eventual "differential" circuit breaker (or connect measurement before it).
5) connect AC voltmeter between GND and Neutral (on the socket).
6) ADVISE EVERY PEOPLE INSIDE YOUR BUILDING!!
The following operation is RISK of FOLGORATION !

7) connetc a power load (like a 100W lamp) between Line and GND. Lamp should turn on, read the ACVoltmeter (V1) and DISCONNECT SOON!

Cool Applying the Ohm law, the GND resistance is Rgnd=V1/(Power_lamp/V_mains)
eg:
V1= 6V
Power_lamp=100W
V_mains=220V
Rgnd= 6/(100/220)=8.2 Ohm, that is good for most applications.
Back to top
Google
AdSense
Google Adsense




Post09 Aug 2004 17:21   

Ads




Back to top
brmadhukar



Joined: 21 Jun 2002
Posts: 844
Helped: 29


Post17 Aug 2004 1:44   

how to test an earth connection


Hi,
check the earth(absolute) to neutral voltage if possible. This should be less than 10V. Also check the votages with respect to neutral. This should not be more than 10V from the pecification.
Make sure the chassis is connected to the ground/neutral.

B R M
Back to top
Arabic versionBulgarian versionCatalan versionCzech versionDanish versionGerman versionGreek versionEnglish versionSpanish versionFinnish versionFrench versionHindi versionCroatian versionIndonesian versionItalian versionHebrew versionJapanese versionKorean versionLithuanian versionLatvian versionDutch versionNorwegian versionPolish versionPortuguese versionRomanian versionRussian versionSlovak versionSlovenian versionSerbian versionSwedish versionTagalog versionUkrainian versionVietnamese versionChinese version
Post new topic  Reply to topic    EDAboard.com Forum Index -> Electronic Elementary Questions -> How to test quality of earth connection ?
Page 1 of 1 All times are GMT + 1 Hour
Similar topics:
battery and connection to earth (4)
Pin connection test in programmers. (1)
How can i detect earth fault? (4)
How to measure resistance between a point and earth? (14)
VOIP Voice Quality - How to qualify? (4)
how to improve mobile reception quality (8)
how to design a high quality LPF (4)
How to calculate the quality factor of HPF (3)
how to simulate quality factor of an spiral inductor in ADS? (2)
how to enhance the Quality factor of a Gm-C elliptic LPF (3)


Abuse || Administrator || Moderators || Support us || sitemap
topic RSS