Pulse Oximeter


Post new topic  Reply to topic    EDAboard.com Forum Index -> Hobby Circuits and Small Projects Problems -> Pulse Oximeter
Author Message
nareshbhar



Joined: 16 Jan 2009
Posts: 10


Post16 Mar 2009 19:23   

ecg pulse oximeter multiplexer


Hi all,
Actually i am making pulse oximeter as my project.
I am using OPT101 as RED & IR detector, and output of the detector is connected to instrumentation amplifier AD630 as shown in schematic and rest of circuit is connected as shown in schematic.CD 4053 is analog multiplexer which is used to separate out RED & IR detector output signal.
but i am not getting expected result at output point(e.g. Point D,E,F,G),so pls anyone could figure out the problem with circuit,specifically filters.
please give your views,& if any alternate circuits.
Thank You.



Sorry, but you need login in to view this attachment

Back to top
Audioguru



Joined: 19 Jan 2008
Posts: 1148
Helped: 95
Location: Toronto area of Canada


Post17 Mar 2009 5:49   

pulse oximeter canada


The inputs of the AD620 need a resistance to ground through a patient. The patient needs a negative common-mode feed usually to the right leg which is near the ground voltage. Then the inputs of the AD620 will work.
Back to top
nareshbhar



Joined: 16 Jan 2009
Posts: 10


Post17 Mar 2009 17:46   

shematic cd4053


but my sensor is finger sensor,and it has nothing to do with direct body contacts.
it just transmits the red & ir light which passes through finger and then received by detector.
Back to top
Audioguru



Joined: 19 Jan 2008
Posts: 1148
Helped: 95
Location: Toronto area of Canada


Post17 Mar 2009 23:01   

gnd an ad620


You showed an ECG circuit.
When signal C1 goes high then the ECG output from the instrumentation amplifier is routed to the upper lowpass filter.
When signal C1 goes low then the ECG signal is routed to the lower lowpass filter.

There is no IR nor red detector circuit.
Back to top
nareshbhar



Joined: 16 Jan 2009
Posts: 10


Post18 Mar 2009 18:20   

Re: Pulse Oximeter


how can you say that its ECG circuit?. i think one can not say anything about circuit's application by just watching schematic.
and about analog multiplexer, its used to differentiate red & ir detector output,which is coming from one light detector only.
Back to top
Audioguru



Joined: 19 Jan 2008
Posts: 1148
Helped: 95
Location: Toronto area of Canada


Post18 Mar 2009 21:41   

Pulse Oximeter


Your ECG circuit does not have inputs for an IR detector and a Red detector. Instead it has an instrumentatation amplifier (look on the datasheet of the AD620 for an ECG circuit similar to yours) that has its output switched to 2 identical filters.

Only 1 of the 3 switches in the multiplexer is switched.
Back to top
nareshbhar



Joined: 16 Jan 2009
Posts: 10


Post29 Mar 2009 14:08   

Re: Pulse Oximeter


yes,actually i have taken this circuit from datasheet only,and as i need two differnt filter so as i can get two outputs,one bieng red detector and one being ir detector.
actually problem with the circuit is that it is not filtering out my desired signal which is around .6hz to 1.5hz.so,can you suggest any change in filter design.
Thanks.
Back to top
Audioguru



Joined: 19 Jan 2008
Posts: 1148
Helped: 95
Location: Toronto area of Canada


Post29 Mar 2009 15:53   

Pulse Oximeter


Your schematic does not show an IR detector and a red detector. It just shows part of an ECG circuit with some filters.
The lowpass filters pass frequencies below about 17Hz.
Back to top
nareshbhar



Joined: 16 Jan 2009
Posts: 10


Post29 Mar 2009 19:00   

Re: Pulse Oximeter


yes, I havn't shown the red & ir detector circuit.
I am using one detector with detecting range of 550nm-850nm.
I have connected my detector output to the input of instrumentation amplifier.
Back to top
Google
AdSense
Google Adsense




Post29 Mar 2009 19:00   

Ads




Back to top
ronydc



Joined: 17 Nov 2005
Posts: 148


Post30 Mar 2009 20:47   

Pulse Oximeter


WHY U EXPECT RESULTS HERE WITHOUT SHOWING THE red and ir DETECTOR CIRCUITS HERE .?

THINK..
Back to top
ostik



Joined: 18 Apr 2009
Posts: 1


Post18 Apr 2009 18:38   

Pulse Oximeter


hi im also making a pulse oximeter, if anyone can pls help me with the software pls. upload an example of a program so i know what its expected to look like atleast. im very new 2 this whole design thing n this is my first project. a schemetic would also be nice
Back to top
Arabic versionBulgarian versionCatalan versionCzech versionDanish versionGerman versionGreek versionEnglish versionSpanish versionFinnish versionFrench versionHindi versionCroatian versionIndonesian versionItalian versionHebrew versionJapanese versionKorean versionLithuanian versionLatvian versionDutch versionNorwegian versionPolish versionPortuguese versionRomanian versionRussian versionSlovak versionSlovenian versionSerbian versionSwedish versionTagalog versionUkrainian versionVietnamese versionChinese version
Post new topic  Reply to topic    EDAboard.com Forum Index -> Hobby Circuits and Small Projects Problems -> Pulse Oximeter
Page 1 of 1 All times are GMT + 1 Hour
Similar topics:
Photoplethysmograph - Pulse Oximeter (2)
BUILDING A PULSE OXIMETER (1)
wavelength for pulse oximeter (8)
Wanted pulse oximeter plans (2)
LM358 as a comparator (Pulse-oximeter) (6)
DESIGNING A PULSE OXIMETER TSL230R BASED (34)
Wireless pulse oximeter design using MSP430 (1)
need help with working pulse oximeter circuit (3)
How to delay a 0.5ms pulse to a 3 second pulse using ladder? (1)
Convert a train of pulse into a single pulse (3)


Abuse || Administrator || Moderators || Support us || sitemap
topic RSS